Enmet EX-5155 Manual do Utilizador Página 14

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 19
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 13
EX-5155-MOS
ENMET Corporation
12
Normal Display Mode
5ppm
MENU
H
OLD the magnet over the
M
ENU switch for 2 – 4 seconds
to enter the Maintenance Menus
Exit
SELECT
Tap the
S
ELECT switch to return to the Normal
Display Mode. See Section 5.2.3
MENU
Tap the magnet over the MENU switch to cycle through Maintenance Menus
OR
Zero
Cal OK
If
the Zero signal is within Preset Specs the EX-5155-MOS will
display Cal OK, See Section 5.2.1
If the Zero signal is not within Preset Specs the EX-5155-
MOS will
display Bad ZERO
NOTE: Some software revisions require the SELECT switch be
tapped to accept the signal.
SELECT
MENU
Bad ZERO
PV: 0
SELECT
Tap the SELECT switch to initiate Zero adjustment
Apply
Calibration Gas until signal value becomes
stable (about 1 to 4 minutes) See Figure 4
If cal is good display will indicate OK or Same
If cal is not within preset “range” display will
indicate Bad Sens
NOTE: Some software revisions require the SELECT switch
be tapped to accept the signal.
MENU
PV: 0
20
Cal OK
Same mV
Bad Sens
Span
SELECT
SELECT
OR
OR
NOTE: You can change the Calibration Gas Level.
HOLD the magnet over the SELECT switch
The MENU switch changes digit indicated by underscore cursor
The SELECT switch locks underscored digit and moves to next digit
See Section 5.2.2
SELECT
Select
Menu
To change Alarm set points:
Tap Menu switch until Alarm to be changed is displayed
Tap Select switch to display the set point
The M
ENU
switch: changes digit indicated by underscore
cursor
The S
ELECT
switch: locks in the underscored digit and
moves to next digit
If change is not within range display returns to first digit
If change is within range display moves to Set Time Delay
Use M
ENU
and S
ELECT
switches as above to change time
delay. Between 0 and 5 seconds is allowed
If change is within range display moves to next menu
Λ - Indicates increasing alarm
V - Indicates decreasing alarm
SetTDsec
0
Alarm3
SELECT
Λ 50
MENU
SetTD
sec
0
Alarm1
MENU
SELECT
Λ 10
SetTDsec
0
Alarm2
MENU
SELECT
Λ
20
Figure 5: Continued on next page
Vista de página 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19

Comentários a estes Manuais

Sem comentários