SE-5175 ENMET Corporation 8 5.0 Maintenance The SE-5175 maintenance menus that are accessed by pressing the MENU button and entering a valid acce
SE-5175 ENMET Corporation 9 Normal Gas Display FIGURE 6: SE-5175 Maintenance Menu Flow Chart 5ppm MENU Press and HOLD the MENU button for
SE-5175 ENMET Corporation 10 5.1.1 Exit Maintenance Menu Exit maintenance, when the Exit appears on the display. Press the SELECT button to retu
SE-5175 ENMET Corporation 11 5.2.1 Zero Adjust NOTE: Zero function is not used for Oxygen transmitters, proceed to Gas Span. The ZERO function m
SE-5175 ENMET Corporation 12 5.2.2 Gas Span It is recommended that the Zero Function be performed first. Do not perform a calibration unless span
SE-5175 ENMET Corporation 13 5.2.3 Alarm Set Points The SE-5175 alarm set points can be changed within limits. See Table 4 for factory set alar
SE-5175 ENMET Corporation 14 5.4 Sensor Replacement WARNING: Power must be removed from the SE-5175 before this or any internal procedure. Failu
SE-5175 ENMET Corporation 15 6.0 Replacement Parts ENMET replacement part numbers: Description of Part Part Number Sensor – Contact ENMET 6702
SE-5175 ENMET Corporation 16 8.0 WARRANTY ENMET warrants new instruments to be free from defects in workmanship and material under normal use for
PO Box 979 680 Fairfield Court Ann Arbor, Michigan 48106-0979 734.761.1270 Fax 734.761.3220 Returning an Instrument for Repair ENMET instrument
Table of Contents 1.0 INTRODUCTION...
Repair Return Form Mailing Address: ENMET Corporation PO Box 979 Ann Arbor, Michigan 48106 Phone Number: 734.761.1270 FAX Number: 734.
SE-5175 ENMET Corporation 1 1.0 Introduction The SE-5175 sensor/transmitter (S/T) instrument is a 3 wire, 24 VDC, 4-20mA S/T, that measures and d
SE-5175 ENMET Corporation 2 2.0 Components of the SE-5175 2.1 SE-5175 elements See Figure 1 for location of elements: Feature Description Enclos
SE-5175 ENMET Corporation 3 2.3 Circuit Board Features The Display Panel is attached to the enclosure cover and is released by unscrewing the 4
SE-5175 ENMET Corporation 4 3.0 Installation The SE-5175 sensor/transmitter detects gas only at the sensor location. NOTE: Sensor Location Gas
SE-5175 ENMET Corporation 5 3.2 Wiring the SE-5175 The electrical installation should conform to appropriate electrical codes, such as the Nation
SE-5175 ENMET Corporation 6 4.0 Operation 4.1 Start Up SE-5175 When the SE-5175 is first powered up, it goes through a series of momentary scree
SE-5175 ENMET Corporation 7 4.2 Normal Display Mode When the SE-5175 is installed as described in section 3, and in clean air, the POWER green LE