ENMET Corporation PO Box 979 Ann Arbor, MI 48106-0979 www.enmet.com Manual Part Number 80003-091 MCN-13-006, 10/24/13 E
EX-5185-PID ENMET Corporation 8 5.0 Maintenance of the EX-5185 CAUTION: Do not open the EX-5185 S/T in a classified area. CAUTION: Do Not Attempt
EX-5185-PID ENMET Corporation 9 5.2 Calibration of the EX-5185 Calibration is the process of setting the instrument up to read accurately when exp
EX-5185-PID ENMET Corporation 10 Table 3: Examples of Standard Calibration Gas Gas Range Alarm 1* Alarm 2* Alarm 3* Sensor Part Number Span G
EX-5185-PID ENMET Corporation 11 5.2.1 Zero Adjust A ZERO function should be performed only when the EX-5185 sensor/transmitter is exposed to fres
EX-5185-PID ENMET Corporation 12 Normal Display Mode 5ppm MENU HOLD the magnet over the MENU switch for 2 – 4 seconds to enter the Maintenanc
EX-5185-PID ENMET Corporation 13 Figure 5: EX-5185 Maintenance Menu Flow chart 5.3 Heater Voltage Settings Heater Voltages are necessary for PID
EX-5185-PID ENMET Corporation 14 6.0 Replacement Part Numbers ENMET replacement part numbers: Description Part Number EX-5185-PID p/n 10014-026,
PO Box 979 680 Fairfield Court Ann Arbor, Michigan 48106-0979 734.761.1270 Fax 734.761.3220 Returning an Instrument for Repair ENMET instru
Repair Return Form Mailing Address: ENMET Corporation PO Box 979 Ann Arbor, Michigan 48106 Phone Number: 734.761.1270 FAX Number: 73
EX-5185-PID ENMET Corporation Table of Contents 1.0 INTRODUCTION ...
EX-5185-PID ENMET Corporation 1 1.0 Introduction The ENMET EX-5185 sensor/transmitters (S/T) is a 3-wire, 24 VDC 4-20mA S/T for the detection of
EX-5185-PID ENMET Corporation 2 2.0 Features of the EX-5185 See Figure 1 for location of features: Feature Description Display LCD: Indicates th
EX-5185-PID ENMET Corporation 3 3.0 Installation of the EX-5185 CAUTION: Area must be declassified during installation. The ENMET EX-5185 gas sen
EX-5185-PID ENMET Corporation 4 3.2 Wiring the EX-5185 to a Control Unit CAUTION: Area must be declassified during installation. Run conduit and
EX-5185-PID ENMET Corporation 5 Figure 3: Terminal Positions EX-5185 Sensor/Transmitter When wiring is complete, reassemble the EX-5185.
EX-5185-PID ENMET Corporation 6 4.0 Operation of the EX-5185 It is best to have the EX-5185 transmitters powered up and operational for 24 hours b
EX-5185-PID ENMET Corporation 7 4.2 Normal Display Mode When the EX-5185 is installed as described in section 3, and in clean air, the POWER gree