Enmet EX-6175 Manual do Utilizador Página 24

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 30
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 23
EX-6175
ENMET Corporation
21
11.2.21 E : 31 – Warm-up timer
The sensor goes through a stabilization period when power is first applied. This option allows the user to adjust the time that is
applied before readings are displayed / transmitted.
From the menu system select menu option: E: 31 and press ENTER.
Using the INCREASE and DECREASE buttons, set the required warm-up time.
Press ENTER to store the new value.
Note: Pressing the MENU button rather than the ENTER button exits without any change.
11.2.22 E : 47 – Analogue Output FSD
This menu option allows the full-scale gas level to be selected that relates to the 4 to 20 mA output.
From the menu system select menu option: E: 47 and press ENTER.
Using the INCREASE and DECREASE buttons, adjust the FSD to the required level.
Press ENTER, “----“ will be displays to indicate that the new setting has been accepted.
Note: Pressing MENU instead of ENTER will exit the menu without changing the FSD.
Press the MENU button to close the menu system
Normally warm time is 45 seconds.
Vista de página 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 30

Comentários a estes Manuais

Sem comentários