Enmet CP-60 Manual do Utilizador Página 1

Consulte online ou descarregue Manual do Utilizador para Instrumentos de medida Enmet CP-60. ENMET CP-60 User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - AINTENANCE

ENMET Corporation PO Box 979 Ann Arbor, MI 48106-0979 Manual Part Number 80003-501 MCN-423, 07/15/09 CP-60 OPE

Página 2 - Table of Contents

CP-60 ENMET Corporation 7 3.4.2 Relay Contacts Auxiliary alarms should be powered from an independent power source separate form the instrument p

Página 3 - List of Tables

CP-60 ENMET Corporation 8 4.0 Operation 4.1 Normal Operation Condition With the CP-60 installed as described in Section 3, and in clean air, the P

Página 4 - 1.0 Introduction

CP-60 ENMET Corporation 9 4.5 Display In clean air a display is shown in Figure 4. This position of the display is termed the "operational d

Página 5 - 2.0 Instrument Features

CP-60 ENMET Corporation 10 5.0 Maintenance The CP-60 has no specific preventative maintenance requirements. Entering the maintenance menu, as out

Página 6

CP-60 ENMET Corporation 11 5.2.1 Set 4 –20mA Transmitter Scale This section of the maintenance menu is for external sensor/transmitters. This fu

Página 7 - 3.0 Installation

CP-60 ENMET Corporation 12 5.2.2 Set Alarm Points Factory alarm set points are discussed in Section 4.2, See Table 1. To change the alarm points

Página 8 - Recommended Wire Gauge

CP-60 ENMET Corporation 13 Example: Set Alarms Flow Chart Displays shown are factory default settings. Λ - Indicates alarm triggered on increasing

Página 9 - 3.4.1 Outputs

CP-60 ENMET Corporation 14 5.2.3 Set Alarm “Delay On” The alarms may be set to delay activation by 1 second increments, from 1 to 30 seconds. WARN

Página 10 - 3.4.4 Digital Outputs

CP-60 ENMET Corporation 15 5.2.5 Failsafe Configuration The CP-60 is factory set in a failsafe configuration. This means that if power is discon

Página 11 - 4.0 Operation

CP-60 ENMET Corporation 16 5.2.7 Set New Password To change the password, you must enter the maintenance menu. Press the OPTION switch until “Ent

Página 12

Table of Contents 1.0 Introduction ...

Página 13 - 5.0 Maintenance

CP-60 ENMET Corporation 17 6.0 Technical Data and Specifications The technical data and specifications of the CP-60 instrument. Electrical Power 1

Página 14

CP-60 ENMET Corporation 18 8.0 WARRANTY ENMET warrants new instruments to be free from defects in workmanship and material under normal use for a

Página 15 - 5.2.2 Set Alarm Points

PO Box 979 680 Fairfield Court Ann Arbor, Michigan 48106-0979 734.761.1270 Fax 734.761.3220 Returning an Instrument for Repair ENMET instrumen

Página 16

Repair Return Form Mailing Address: ENMET Corporation PO Box 979 Ann Arbor, Michigan 48106 Phone Number: 734.761.1270 FAX Number: 734.

Página 17 - 5.2.3 Set Alarm “Delay On”

List of Figures Figure 1: CP-60 External Features ...

Página 18 - 5.2.6 Set Output Span Range

CP-60 ENMET Corporation 1 1.0 Introduction The CP-60 control panel is capable of monitoring from 1 to 3 remote sensors. Some features of the moni

Página 19 - 5.2.8 Exit Maintenance Menu

CP-60 ENMET Corporation 2 2.0 Instrument Features 2.1 CP-60 Exterior Features The exterior of the instrument is shown in Figure 1. The exterior f

Página 20 - 7.0 Replacement Part Numbers

CP-60 ENMET Corporation 3 2.3 Circuit Board Features The Display Panel is hinged on the left and is released by unscrewing the 2 philips screws lo

Página 21 - 8.0 WARRANTY

CP-60 ENMET Corporation 4 3.0 Installation 3.1 Mounting of Instrument Mount the CP-60 instrument on an appropriate vertical surface using the moun

Página 22

CP-60 ENMET Corporation 5 3.4 Sensor/Transmitter Connection Sensor/Transmitters are connected to the CP-60 control unit with two or three-conducto

Página 23 - Repair Return Form

CP-60 ENMET Corporation 6 3.4.1 Outputs Two types of alarm outputs are available, relay contacts and 4-20mA outputs. Figure 5: CP-60 Relay, Inpu