Enmet SPECTRUM Manual do Utilizador Página 24

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 32
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 23
ENMET Corporation SPECTRUM Series
20
SPECTRUM Series for Nitric Oxide (3E 100 sensor)
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor Concentration in ppm
chlorine 5
nitrogen dioxide 100
hydrogen 1000
carbon monoxide 1000
carbon dioxide 10000
saturated HC, alcohols abundant
sulfur dioxide 50
SPECTRUM Series for Nitrogen Dioxide (3E sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
chlorine 1 3
ozone 1 0.7
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor Concentration in ppm
hydrogen chloride 5
nitrogen oxide 100
hydrogen 1000
carbon monoxide 1000
carbon dioxide 10000
saturated HC, alcohols abundant
sulfur dioxide 50
ammonia 30
SPECTRUM Series for Hydrazine (2E sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
ammonia 200 0,4
methyl hydrazine 1 0,72
1.1 dimethyl hydrazine 1 0,45
chlorine 3 –0,02
hydrogen sulfice 20 0,1
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor Concentration in ppm
ethanol 1000
Isopropanol 450
hydrogen peroxide 10
hydrogen chloride 7
methanol 1200
hydrogen 2000
carbon monoxide 1000
carbon dioxide 5000
unsaturated HC (ethylene) 1%