Sensor/Transmitter, Toxic Gas & Oxygen ENMET Corporation
12
Fluorine (3E 10 sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
chlorine
1
2
hydrogen sulfide
10
-
1
sulfur dioxide
5
-
3.2
nitrogen dioxide
5
.5
bromine
1
3
ammonia
1000
-
3.1
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
carbon monoxide
300
carbon dioxide
100000
nitrogen
100%
hydrogen chloride
10
hydrocarbons
% range
hydrogen
1000
hydrogen cyanide
10
ethanol
4%
Hydrogen (2E 2000 sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
carbon monoxide
50
0
ethylene
500
0
sulfur dioxide
2
0
isopropanol
1090
180
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
ammonia
100
carbon dioxide
1000
chlorine
5
hydrogen cyanide
10
hydrogen sulfide
10*
methane
10000
nitrogen dioxide
10
sulfur
dioxide
2
*With onboard filter; continuous high level exposure may reduce the filter efficiency