Enmet OX-2000 Manual do Utilizador Página 14

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 24
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 13
ENMET Corporation TX2000 / OX2000
10
4.3 Calibration
Verify that the TX/OX 2000 is working properly by calibration, performed with the help of a calibration gas. If
the response is not adequate or when any of the following occur, perform a complete calibration.
At least every 6 months for the
Whenever the instrument has been exposed to high concentrations of gas.
When an instrument has been stored for more than one month without use.
When changing of sensors.
If “DEF16” fault appears (measurement too negative).
Within the calibration menu, "visible and audio" alarms are not triggered.
“Calibration” menu consists of:
Setting the zero in clean air and the sensitivity with calibration gas. See Section 4.3.1
There is no zero adjustment for the OX2000. See section 4.3.2
4.3.1 TX2000 Calibration
Enter maintenance menu.
Press both switches at the same time for 5 seconds:
&
(During this time, measurement is displaying)
5 seconds later, the maintenance pictogram appears
Release the switches
At this time, the operator has 5 seconds to enter the
maintenance menus or the instrument returns to operation
mode.
· Then proceed with following
Press BACKLIGHT switch to toggle
between (PrG) and (CAL)
to continue See Sections:
4.2 for Programming
4.3.1A for Calibration
Figure 4: Entrance to Maintenance Menu Flow Chart
Do you want to reach
the menus?
N
O
YES
No action on switches
Return to normal operation,
5 seconds late
r
Press ALARM switch 3 times
Maintenance pictogram is flashing
You have access to:
Calibration (CAL) and
Programming (PrG) menus.
Vista de página 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 23 24

Comentários a estes Manuais

Sem comentários