Enmet MRI-5175 Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Instrumentos de medida Enmet MRI-5175. ENMET MRI-5175 User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 2
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 0
ENMET Corporation
80003-503
Original April 2011
Page 1 of 2
Page 1 of 2
Guide for MRI-5175 Sensor Replacement
An Oxygen sensor must be replaced when it can no longer be calibrated. Typical sensor life is 3 to 5 years in a 20.9%
oxygen environment.
To replace a sensor, perform the following steps:
Turn off the electrical power.
Open the display panel and remove the four cover (lid) retention screws and remove the lid. See Figure 1.
Remove the old O
2
sensor, by unscrewing it, and replace it with a new sensor. See Figure 1.
Replace cover(lid) with retention screws.
Turn on the electrical power.
Wait 3 – 4 hours, then recalibrate the sensor per the instructions.
MRI-5175 Exterior View
(Optional)
V+
SIG
GND
MRI-5175 Lid, Interior View
Sensor Wiring
J8 Sensor Wires
V+
Not Used
Sig
Red wire
Gnd
Black wire
Figure 1: MRI-5175 Sensor Replacement
Figure 2: Calibration MRI-5175
O2 Sensor
Sensor Wires
To J8
Terminal J8
Power/Fault LED
MENU Switch
SELECT Sw
itch
Calibration Adapter
on Sensor
Menu
Select
Power/
Fault XX X
XXX
XXX
% O2 b y vol.
Cylinder,
Calibration Gas
Vista de página 0
1 2

Resumo do Conteúdo

Página 1 - Sensor Wiring

ENMET Corporation 80003-503 Original April 2011 Page 1 of 2 Page 1 of 2 Guide for MRI-5175 Sensor Replacement An Oxygen sensor must be replaced when

Página 2 - Altitude Calibration

ENMET Corporation 80003-503 Original April 2011 Page 2 of 2 Page 2 of 2 Standard Calibration The following procedure is suitable for altitudes below

Comentários a estes Manuais

Sem comentários