Resumo do Conteúdo

Página 1 - OXYGEN / TOXIC

ENMET Corporation PO Box 979 Ann Arbor, MI 48106-0979 www.enmet.com Manual Part Number 80003-124 April, 2013

Página 2 - Table of Contents

EX-6175 ENMET Corporation 7 3.8 Initial Gas Testing EX-6175 units are factory calibrated as detailed on the calibration certificate supplied with th

Página 3 - List of Figures and Tables

EX-6175 ENMET Corporation 8 4.0 Operation The EX-6175 incorporates magnetically operated switches to allow non-intrusive operation of the menu syste

Página 4 - 1.0 Introduction

EX-6175 ENMET Corporation 9 5.0 Calibration 5.1 Sensor Calibration This feature allows the sensor to be calibrated. Ensure that the correct sensor

Página 5 - 2.0 Features of the EX-6175

EX-6175 ENMET Corporation 10 6 Sensor Replacement WARNING: Power must be removed from the EX-6175 before this or any internal procedure. Failure to

Página 6 - Figure 2: EX-6175 Mounting

EX-6175 ENMET Corporation 11 Step 3 Remove display board which can be left to dangle on its retaining wire. Remove PCBs from the support standoffs

Página 7 - Mount vertically

EX-6175 ENMET Corporation 12 Step 5 Remove sensor insert form EX-6175 enclosure using extraction tool. Step 6 Separate the two sensor insert comp

Página 8

EX-6175 ENMET Corporation 13 Step 7 Apply Loctite 243 or an equivalent to the threads; insert the sensor housing into the EX-6175 enclosure using t

Página 9 - 3.7 Indicators

EX-6175 ENMET Corporation 14 Step 9 Reattach cable, the boards removed form the support standoff and finally the display board back into position.

Página 10 - 3.8 Initial Gas Testing

EX-6175 ENMET Corporation 15 7 Fuses A 1Amp anti-surge fuse, housed within a surface mounted holder, is located within the EX-6175 to prevent over

Página 11 - 4.0 Operation

EX-6175 ENMET Corporation 16 9 Certification The EX-6175 (FGD10 Series) of Gas Detectors are ATEX and IECEx certified for use in potentially explos

Página 12 - 5.0 Calibration

Table of Contents 1.0 INTRODUCTION ...

Página 13 - 6 Sensor Replacement

EX-6175 ENMET Corporation 17 11 Menu Mode The EX-6175 uses a password system to restrict the end user from carrying out certain changes that may com

Página 14

EX-6175 ENMET Corporation 18 11.2.6 E : 6 – Decimal Places The number of decimal places that are used to display the detected gas levels can be chan

Página 15 - Extraction Tool

EX-6175 ENMET Corporation 19 11.2.11 E : 13 – Relay 1 The unit is fitted with a relay that is operated in conjunction with the alarm level. The user

Página 16

EX-6175 ENMET Corporation 20 11.2.16 E : 20 – Negative Zero Suppression This option is used to allow the user to suppress small amounts of negative

Página 17

EX-6175 ENMET Corporation 21 11.2.21 E : 31 – Warm-up timer The sensor goes through a stabilization period when power is first applied. This option

Página 18 - 7 Fuses

EX-6175 ENMET Corporation 22 Appendix A A.1 Analog Output Calibration The analogue output is that of a current source. In order to calibrate the out

Página 19 - 9 Certification

EX-6175 ENMET Corporation 23 A.2 Communications Protocol The communications protocol used by the EX-6175 is used for communications between devices

Página 20 - 11 Menu Mode

EX-6175 ENMET Corporation 24 A.3.2 Reading a Variable Send a read frame with the Variable ID to be read: DLE RD var-id DLE EOF Csum hi Csum lo

Página 21

EX-6175 ENMET Corporation 25 WARRANTY ENMET warrants new instruments to be free from defects in workmanship and material under normal use for a pe

Página 22 - 11.2.12 E : 14 – Relay 2

PO Box 979 680 Fairfield Court Ann Arbor, Michigan 48106-0979 734.761.1270 Fax 734.761.3220 Returning an Instrument for Repair ENMET instrumen

Página 23

2 List of Figures and Tables FIGURE 1: EX-6175 FEATURES ...

Página 24

Repair Return Form Mailing Address: ENMET Corporation PO Box 979 Ann Arbor, Michigan 48106 Phone Number: 734.761.1270 FAX Number: 734.

Página 25 - Appendix A

ENMET Corporation EX-6175 1 1.0 Introduction The ENMET EX-6175 Fixed Gas Detectors for the detection of Flammable gases is ATEX Flameproof Certifie

Página 26 - A.3 Frame Structure

EX-6175 ENMET Corporation 2 2.0 Features of the EX-6175 The EX-6175 is an ATEX and IECEx certified fixed gas detector for use in potentially explos

Página 27 - A.4. Writing a Variable

EX-6175 ENMET Corporation 3 3.0 Installation of the EX-6175 CAUTION: Area must be declassified during installation. The EX-6175 gas sensor/transmit

Página 28 - Notes:

EX-6175 ENMET Corporation 4 Important Connect enclosure earth-ground to a high integrity earth ground ATEX certified Ex d armoured cable glands -

Página 29

EX-6175 ENMET Corporation 5 Figure 5: Unplugging the Display Board for Connection Purposes 3.2 Power Supply The EX-6175 is powered from an 8 – 24 v

Página 30 - Repair Return Form

EX-6175 ENMET Corporation 6 3.5 Following Completion of the Installation: a) Replace the top display board using the locating pins on the underside